|
[Sequential]
[Segmental]
|
As described earlier, spin echo sequences utilise a 180
RF
pulse to rephase the dephasing spins and produce a detectable RF signal. There
are three main types of spin echo sequence; single echo, multi-echo and
echo-train spin echo. An example of a single echo sequence is shown in
figure 2.4. RF excitation and slice selection are
followed by phase encoding of the signal. The 180
rephasing
pulse and slice selection gradient are applied and the echo is
read-out at time
. The sequence is repeated using a different
phase-encoding magnitude each repetition until k-space is filled, as
shown in fig. 2.5(a).
Multiple slices can be acquired during one
period using a
multislice loop structure.
Single echo sequences are typically used to produce
-weighted
images, using a short
(
700ms) and short
(
30ms) [5].
Multi-echo sequences use multiple 180
pulses to produce signals
at different
values, where the only differences between the
signals at the two echo times will be due to the relaxation rates of
the tissues. Multi-echo sequences are typically used to produce two
simultaneous images: a proton-density weighted image where
![]()
30ms and a
-weighted image with ![]()
80ms [5].
is
sufficiently great to allow (almost) complete
relaxation.
Echo-train spin echo sequences are similar to the multi-echo
sequences, except that each spin echo is acquired with a different
phase encoding
as well as its own
. The echo-train length (or turbo factor) is the
number of echos acquired in each
period. Here, k-space is filled
segmentally (see figure 2.5(a)), with one echo from each
echo-train filling each segment of k-space, and as such is an
efficient sequence. These sequences are
typically used to create
-weighted images, with a typical
echo-train of 9 echos used in brain imaging.